Coronary
Collection of technologies in pursuit of trackability
― Kamui XS is a new option in challenging situations ―
■Tapered Outer Tube
The structure of the Outer Tube being pulled into the balloon smoothes the rigidity gap between the balloon and the Outer Tube.
Since the Outer Tube tip is welded to the Inner Tube, a new Inflation Lumen is provided.
■Tapered Hypotube & Cobalt Chromium Core
The high tension Hypotube is processed by centerless machining used in guidewire development, resulting in a tapered design.
The highly rigid Cobalt Chromium Core is used to transmit the pushing force to the tip without waste.
■Inner Coil
By wrapping the flat wire coil around the Inner Tube, the lumen is maintained even in tortuous areas of the vessel, thereby reducing the resistance of the guidewire.
■Push Ring
When the Push Ring hits the GW port, it pushes the entire Distal shaft forward, increasing pushability.
■Profile
The reduced size of the profile compared to the conventional product allows for better penetration into lesions.
■Rigidity Transition
The new construction, in which the Outer Tube is drawn all the way into the balloon, provides smoother rigidity than conventional products.
Hypotube also has a higher rigidity than conventional material, and this new structure overcomes the rigid gap of the catheter.
The above data was (All the data were) obtained by company standardized test, which may differ from industry standardized tests.
The above data (All the data) does not prove that all devices have exactly the same performance with the samples used for these tests.
Balloon Diameter | 0.88mm |
---|---|
Balloon Length | 5mm |
Nominal Pressure | 6 atm |
Rated Burst Pressure | 14atm |
Usable Length | 145cm |
Structure | |